Control Solution Testing:
You should perform a command test: ▪ When using your meter for the first time ▪ To check if y'all are testing correctly |
|
▪ When you open a new vial of test strips
▪ If you get out the exam strip vial open for an extended period of time
▪ If you think your meter may not be working properly
▪ If your test results do not match how you experience Low, Normal and Loftier command solutions
are bachelor.
Utilize but Profile® command solutions (Normal, Low, or High) to practice on
the system or to run a cheque to make sure the meter and the exam strips are working
properly. Using anything other than CONTOUR® command solution could nowadays
incorrect results. Remove a test strip from the bottle. Tightly close the bottle lid
immediately later on y'all take removed the test strip.
Note: Check the expiration and discard dates. Brand sure the test strip does
not announced torn or damaged.
Concord the test strip with the grayness end facing upwards. Insert the grayness cease into the exam strip port on the meter. |
|
The meter volition turn on. An epitome of a examination strip with a flashing blood drib will appear letting y'all know the meter is ready to examination. |
|
NOTE: Gently rock the control canteen before opening to ensure information technology is mixed well. | |
Squeeze a small drop of control solution on a clean non-absorbant surface (such as a piece of wax paper). Practise not use control solution to the examination strip straight from the canteen. |
|
Immediately affect the tip of the test strip to the drop of control solution. The solution is pulled into the test strip through the tip. Hold information technology in the drop until the meter beeps. | |
After a beep you will run across the meter count down 5 seconds until the test is complete and your control test result is displayed. |
|
The meter volition automatically recognize and mark the command result for you. NOTE: The does not indicate the control test result is within the speci- fied range. It only marks it as a control solution test. It will not exist included in your averages. |
|
Compare your control examination result with the normal control range printed on the test strip bottle label or on the bottom of the examination strip box. Notation: Low and High control range values can be establish on the bottom of the test strip box. |
|
To plow your meter off, simply remove the test strip. The consequence will be automatically marked and stored in the meter retention. Control results will not be included in your claret glucose averages or in the 7 day HI-LO summary. Carefully dispose of the used examination strip. |
|
Important: To ensure accuracy of your claret glucose test results and accuracy of your control results, use simply CONTOUR® control solution. If your control test is out of the range listed on the bottle, please contact Customer Back up at 1-800-348-8100. |